mission® hb hemoglobin testing system

quick

• hemoglobin (hb) and hematocrit (hct) results in < 15 seconds

• simply insert strip, apply specimen and read results

• low maintenance and quick clean up

accurate

• precise results equivalent to automatic hematology analyzers

• wide hb measurement range of 4.0-25.6 g/dl with hct range of 11-75%

• hb total precision cv3%

convenient

• only 10 μl capillary or venous blood required

• minimal training required, no specimen preparation

• easy-to-read, large lcd

• quick data transfer via mini usb port




product parameters

feature

specifications


methodology

reflectance photometry


detection  principle

methemoglobin


time to results

< 15 seconds


memory

1,000 tests with date/time and id number


specimen volume

10 μ‑l


specimen type

capillary and venous whole blood


hb measurement  range

13-75%


hct range

525 nm


wavelengths

mini usb port


pc interface

252mm×255mm×151mm


calibration

automatic


hb within run  precision cv

3%


hb total  precision cv

3%


accuracy

venous blood: y=0.9582x 0.5673, r2=0.992;
  capillary blood: y=1.006x 0.026, r2=0.993.


operating  conditions

10-40ºc (50-104ºf); 90% rh


meter storage  conditions

0-50°c (32-122°f); 90% rh


strip storage  conditions

2-30°c (36-86°f); 85% rh


strip shelf life

2 years unopened canister; 3 months opened canister


control solution  shelf life

6 months unopened bottle; 30 days opened bottle


power source

3 aaa batteries or ac adaptor


battery life

2,700 tests or 360 hours


automatic shut  off

8 minutes


meter dimensions  (l x w x h)

127 mm × 58 mm × 25 mm (5.0 x 2.28 x 0.09)


lcd dimensions  (l x w)

39 mm × 37 mm (1.54 x 1.46)